Impedance Measurements — A guide to measuring impedance with Moku:Lab's Frequency Response Analyzer (Part 1)

In this application note, we present some examples of making accurate impedance measurements with Moku:Lab. In this first part, we will explore the maths of measuring resistance and accuracy. In the second part, we expand to explore capacitive and inductive measurements. 

Download Printable Version

Frequency Response Analyzer

Moku:Lab’s Frequency Response Analyzer (FRA) drives a swept sine wave on the Moku outputs and simultaneously measures the received signal amplitude (or power) on the Moku input. This can be used to measure the transfer function of a system or (device under test) and thus create a plot of amplitude and phase vs frequency, commonly referred to as a Bode plot.


Power units

In order to measure impedance of a device under test (Zdut), we need to understand the power plots of the FRA. The FRA plots use units of dBm, or decibels relative to one milliwatt (1mW); a convenient unit of measure in this situation. Defined as:

Moku:Lab’s FRA swept sine output can be set in volts (peak to peak) or dBm. We will use volts for the output. By definition for a sinusoid,

substituting for Vrms in (2) gives

So 1 Vvpp expressed in dBm, scaling for mW and knowing the Moku:Lab input impedance is 50Ω gives:

Figure 1 shows use of the FRA to generate a 1 Vvpp sine wave and with Moku:Lab’s output 1 connected via direct coax to input 1. The resulting amplitude is of course flat across frequency range (0-1kHz) and at 4.050dBm, very close to the calculated 3.979dBm. The discrepancy equates to 1.7mV (or 0.17%).

 

Figure 1: FRA plot of 1 Vvpp driven directly in Moku:Lab’s input

 

Impedance

Single port measurement:
Now that the FRA power units are understood we can measure an impedance.

In this first example we are going to measure a Rdut of a simple 10k ohm, 10% tolerance resistor. The effective circuit is:

 

Figure 2: Equivalent circuit

 

Note, Vout is 2v, since this results in 1V across load of 50Ω.

Moku:Lab’s FRA operates up to 120MHz, but for these resistance measurements, a plot to 40kHz is sufficient. Figure 3 shows Moku:Lab’s FRA amplitude response = -35.821 dBm at Vin.

 

Figure 3: FRA of 10kΩ DUT, single port

 

Rearranging power equation (1) and substituting P from (4) we can state:

With measured Pdb of -35.821dBm, we calculate Vin= 10.23mV.

The resistor divider of Rdut and Moku:Lab’s 50Ω inputs and output in figure 2 gives us:

Thus

Solving gives Rdut = 9675 ohms.

A DVM reading of this resistor showed 9750 ohms.

So from this simple, one resistor measurement we can conclude Moku:Lab is accurate within 77 ohms (< 1%).

Low impedance measurement:
The example above used a standard 10% tolerance resistor. We are now interested in measuring a lower impedance to a high level of accuracy. To do this we will use a 100 ohm, 0.005% tolerance high precision resistor. Using the above method we obtain a power magnitude plot (Figure 4)

 

Figure 4: FRA of 100Ω, 0.005%, single port

 

Applying the measured power of -1.972 dBm to equations (5) & (7) we calculate Rdut to be 98.41 ohms. This agrees with the known value, but we can do better with a two port measurement.

Two port measurement:
In order to have an improved measurement, we need to account for the loading of the DUT on the Moku’s 50ohm output.

We can accomplish this with a two port measurement whereby we utilize the second input port of the Moku to observe the actual applied signal level.

 

Figure 5: Two port configuration

 
 

Figure 6: Two port equivalent circuit

 

We can derive Rdut in figure 6, from ohms law:

Substitute (9) into (8)

or

We setup this two port measurement with our tight tolerance 100Ω, 0.005% resistor and captured the Moku:Lab’s FRA plot in Figure 7.

 

Figure 7: 100 ohm, two port

 

Note, we have used the Frequency Response Analyzer math channel to produce V2/V1. This is very quick and simple to configure on the iPad interface.

From (10) we see that the Rdut can be calculated simply from the V2:V1 voltage ratio.

The Frequency Response Analyzer math channel has calculated the power ratio as 9.505 dBm and thus the voltage ratio is:

So we can apply (10) to obtain Rdut =99.36 ohms.

We can now apply this 2 port method to the original 10k / 20% resistor; Figure 8 shows the FRA response

 

Figure 8: 10k ohm, 2 port

 

Using our established formula; power ratio of 45.856 dBm gives Rdut = 9762 ohms. A useful improvement.


Summary

The Moku:Lab FRA can be used to make impedance measurements and determine to a <1% accuracy, the value of a resistance.

The two port method allows for the loading of the DUT and appears more accurate.

Now that this method and calculations are explained, in part 2 we extend the examples to using the FRA to measure complex impedances, capacitance & inductance, and to more complex systems and across frequency.

Thanks for reading. You can download the Moku:Lab App from Apple’s App Store and experience it in demo mode. Specific features of Moku:Lab’s FRA can be found at liquidinstruments.com/manuals-and-guides/